ECCOS® Shaver tips

Basic Requirements for ECCOS®
> A single chamber cleaning/disinfecting machine is required
> A suitable cleaner must be used (pH < 11 for Aluminum components)
> The temperature in the cleaning cycle must not exceed 55°C
> Clean for at least 3 minutes
> Apply intermediate rinse for at least 1 minute
> Carry out intensive final rinse with distilled, demineralized or fully desalinated water
> Only Thermal disinfection: 5 minutes rinse at 93°C with distilled, demineralized or fully desalinated water
> Complete the program with a drying phase at least 10 Minutes at a temperature of not more than 120°C